Features:

  • Rectangular doors throughout.  The exterior entry door to the critical care lock is a large, easy to use 39"x 72" (99.1 cm x 182.9 cm) rectangular door, the largest standard door we offer.  All the other doors are 34" x 72" (86.4 cm x 192.9 cm)
  • The door sills are flush with both the chamber interior and exterior building floors, substantially simplifying patient transport through the doors.
  • Thirteen 12.5" (31.8") diameter shell windows and  a 10" diameter (25.4 cm) window in each door allow maximum visibility of chamber interior.
  • Console is simple to operate.  Training in complex computer operations not required.
  • Segmented main PVHO permits installation in locations with limited access.

Chamber Capacity:             

Reimers 12000T Series
Triple-Lock Multiplace Hyperbaric System

The 12000T series of multiplace chambers is designed for clinical hyperbaric therapy applications including those involving saturation decompression exposure. The triple lock design allows easy access into and out of the chamber while the chamber is at depth. The 12000T series is available with seating for 19-31 patients and up to 5 attendants.The entry lock of this chamber includes a sink with hot and cold running water and is, itself, capable of treating critical care patients as well as providing ingress and egress for the other two locks. This Chamber offers lots of room, versatility and provides exceptional patient comfort.




System Performance:






    Treatment Lock 1
    Entry Lock
   Treatment Lock 2
Patients, Seated
      Up to 14
      Up to 3
      Up to 14
Attendants Stations
           2
           1
           2
Diameter, Interior
    120 inches (304.8 cm)
    120 inches (304.8 cm)
   120 inches (304.8 cm)
Useful Interior Length
     18' 0"  (5.48 m)
     8' 1-1/2" (2.47 m)
     17' 6-1/2" (5.34 m)
Headroom, Floor to Ceiling
      7' 4" (223 cm)

     7' 4" (223 cm)
      7' 4" (223 cm)
Pressure Capability
               6.5 ATA (81 psig)
       6.5 ATA (81 psig)

        6.5 ATA (81 psig)
Pressurization Rate
               6 psi/min
           >10 psi/min
           >5 psi/min
Sound Levels
  Treatment conditions (1)
  Pressurization  (2)
  Depressurization  (2)
          62.5 dBA
        < 65 dBA at 10 fsw/min
        < 78 dBA at 10 fsw/min

<70 bBA
< 84 dBAat 15 fws/min.
< 78 dBA at 15 fsw/min

62.5 dBA
< 65 dBA at 10 fsw/min
< 78 dBA at 10 fsw/min

Lock aid condition
Standard
Optional        
Standard
Chamber Specification:



One-Piece Pressure Vessel
Segmented Pressure Vessel
Overall Length
       46' 1" (15.05 m)
46' 1" (15.05 m)
Overall Width
     10' 8" (325.1 cm)
10' 11-1/8" (307.6 cm)
Overall Height
10' 9"  (327.7 cm)    
11' 1" ((337.8 cm)
Required Installation Opening
       11' 6"  high x 11' wide (350 cm x 335cm)
7'  by 11' 1" (213.4cm by 338cm)
Overall Weight (Shell Only)
       106,000 lbs  (48,200 kg)
120,000 lbs (54,500 kg) 
Overall Weight (Outfitted)
       126,000 lbs  (57,300 kg)
140,000 lbs (63,600 kg) 
Hyperbaric System Includes:


Air Pressurization and Ventilation
* Oil-less air compressors (size and model to suit site conditions)
* Automatic profile controls
* Back up manual pressurization and ventilation controls
Breathing Gas
* Patient stations support masks, hoods, and nebulizers
* Attendant stations support masks
* Patient hoods/masks work normally with chamber at surface pressure
* Operates on standard 50 psig hospital oxygen supply pressure
* Automatically switches off oxygen if the fire system is activated
* High-pressure back-up oxygen and breathing air manifolds
Fire Suppression & Potable Water
* Deluge in each lock.  Two hand lines in treatment locks, one in entry lock.
* Deluge and hand line are fully independent systems.
* Bypasses  installed so deluge can be tested without flooding chamber.
* System designed so water stays clean. Potable water via hand line system.
Communications
* Wireless  primary with backup sound powered phone
Atmospheric Conditioning
* Uses building hot and cold water
Gas Analysis
* Full time oxygen analysis each lock with capability to monitor breathing gas
Patient Entertainment
* AM/FM cassette feed into chamber
Potable Water (optional)
* Hand washing sink in main lock for infections control as per JACO req'ts
Video Monitoring
* Provides full time monitoring of Chamber
Medical Suction
* Support for standard medical suction devices provided in all locks
Patient Ventilators
* Support for pneumatically powered patient ventilators provided in all locks
Notes:        
1.      ACS blower operating at 50% of capacity, chamber ventilation rate at 100 scfm (50% of capacity)
2.      From normal control valves.  Values are higher when using emergency manual valves.
3.      Each lock can be pressurized independently of the others.


Reimers Systems, Inc.  All Rights Reserved 2009.

Lease
RSI Products > Hyperbaric Chambers > Model 12000T
Additional options are available for each system, which add features to increase functionality or remove unneeded features to reduce cost.

All Reimers Systems hyperbaric systems are designed to fully meet the current edition of all applicable codes including:

  • NFPA 99; "Health Care Facilities" Chapter 20, "Hyperbaric Chambers"
  • ASME PVHO-1; "Safety Standards for Pressure Vessels for Human Occupancy"
  • ASME Section VIII, Division I, "Unfired Pressure Vessels"