Model: 9612D
|
|
Patients, Seated, Mail Lock
|
12
|
Patients, Seated, Entry Lock
|
4
|
Pateints, Supine, Main Lock
|
4
|
Attendents, Seated, Main Lock
|
4
|
Attendents, Seated, Entry Lock
|
2
|
Pressure Capability
|
6.5 ATA (81 psig)
|
Pressurization Rate
|
6 psi/min
|
Overall Length
|
30' 0" (915 cm)
|
Overall Width
|
8' 1" (216 cm)
|
Overall Height
|
8' 5" (226 cm)
|
Interior Length, Main Lock
|
25' 0" (318cm)
|
Interior Width, Main Lock
|
8' 0" (213 cm)
|
Interior Length, Entry Lock
|
5' 0" (142cm)
|
Interior Width, Entry Lock
|
8' 0" (213cm)
|
Interior Ceiling Height
|
6' 3" (190 cm)
|
Air Pressurization and Ventilation
|
* Air plant with dual oil-less compressors
* Automatic profile controls
* Back up manual pressurization and ventilation controls
|
Breathing Gas
|
* Patient stations support masks, hoods, and nebulizers
* Attendant stations support masks
* Automatically switches off oxygen if the fire system is activated
* High-pressure back-up oxygen and breathing air manifolds
|
Fire Suppression
|
* Deluge and two hand lines
* Deluge and hand line are independent systems
* Bypass installed so deluge can be tested without flooding chamber
|
Communications
|
* 4 station primary with backup sound powered phone
|
Atmospheric Conditioning
|
* Uses building hot and cold water
|
Gas Analysis
|
* Full time oxygen analysis
|
Patient Entertainment
|
* AM/FM cassette feed into chamber
|
Potable Water (optional)
|
* Hand washing sink in main lock for infections control as per JACO req'ts
|
Video Monitoring
|
* Provides full time monitoring of Chamber
|