Model B-12 Hyperbaric Chamber system simulates the hydrostatic pressure cycle encountered by fish passing through a hydroelectric dam. The System is designed to be portable and to be used at a location by the water, where fish can be collected and cycled through the chamber. The system is designed to circulate the water available from the same location that the fish are gathered from.


Overview:

Each chamber module includes water pressure and flow controls, pressure and temperature sensors

Each chamber has the capability to maintain a set "acclimation" pressure indefinitely then transition into the fast pressure change simulation of the fish passing through a turbine

Specification:


Chamber water volume
50 liter
Chamber working pressure
1 to 60 psia with 14.7 psia ambient
Acclimation pressure range
20 to 60 psia with 14.7 psia ambient
Acclimation pressure control
+1 psi
Acclimation press/depress rate
Up to 60 psi/min
Spike pressure range
1 to 60 psia with 14.7 psia ambient
Spike press/depress rate
1 to 300 psi/sec
Access port opening
12-inch diameter.
Acclimation water flow range
1 to 5 gpm (less at 5 psig or less)
Water flow accuracy
+0.5 gpm
Air pocket control
Automatic control
Viewing Features
4 sided flat lens for minimum optical distortion
Fish removal system
4" diameter lower drain.  Manual operated full port ball valve
Spare ports
2 each 1/2" NPT ports
Water temperature range
5 to 30 degrees Celcius
Pressure sensing accuracy
+0.2 psi
Computer control software
National Instruments Labview
Standby Computer
Identical hardware as Primary
Pumping System
2 pumps of 3/4 hp each supplied by a water tower system to maintain a steady suction head.
Compressed Air System
1 air compressor of 3/4 hp. 2.5 cfm @ 100 psi
Reimers Systems, Inc.  All Rights Reserved 2011.

RSI Products > Research Chambers > Model B-12
One-Year Limited Warranty

Reimers Systems, Inc. (RSI) will replace or repair, at its sole option, any part found to be defective due to reasons of materials or workmanship within one year from the date of delivery provided that the affected items are shipped pre-paid to RSI. Return freight will be collected.