Reimers 9600D Series
Dual-Lock Multiplace Hyperbaric System

The 9600D series of multiplace chambers is designed for clinical hyperbaric therapy applications including those involving non-saturation decompression exposure. The dual lock design allows easy access into and out of the chamber while the chamber is at depth. The 9600B series is available with seating for12 patients and up to 4 attendants, with additional space in the entry lock. The entry lock of this chamber is designed for use anytime ingress and egress is necessary. This Chamber offers lots of room and versatility.

Features:

  • Exterior main lock is a large, easy to use 32"x 66" (82 cm x 168 cm) rectangular door, the largest standard door we offer.

  • Other doors are 36" (92cm) round doors

  • Ten 10" diameter windows allow maximum visibility of chamber interior.

  • Upholstered chairs are comfortable and quickly fold up to allow up to four gurneys in the main lock.

  • Attendant has folding jump seats that fold up out of the way.

 

Chamber Capacity:

  9612D
Patients, Seated, Main Lock 12
Patients, Seated, Entry Lock 4
Patients, Supine, Main Lock 4
Attendants, Seated, Main Lock 4
Attendants, Seated, Entry Lock 2

System Performance:

Pressure Capability  6.5 ATA (81 psig)
Pressurization Rate 6 psi/min

Chamber Specifications:

Overall Length 30' 0" (915 cm)
Overall Width 8' 1" (216 cm)
Overall Height 8' 5" (226 cm)
Interior Length, Main Lock 25' 0" (318cm)
Interior Width, Main Lock 8' 0" (213 cm)
Interior Length, Entry Lock 5' 0" (142cm)
Interior Width, Entry Lock 8' 0" (213cm)
Interior Ceiling Height 6' 3" (190 cm)
Overall Weight  

Hyperbaric System Includes:

Air Pressurization and Ventilation Air plant with dual oil-less compressors
  Automatic profile controls
  Back up manual pressurization and ventilation controls
Breathing Gas Patient stations support masks, hoods, and nebulizers
  Attendant stations support masks
  Automatically switches off oxygen if the fire system is activated
  High-pressure back-up oxygen and breathing air manifolds
Fire Suppression Deluge and two hand lines
  Deluge and hand line are independent systems
  Bypass installed so deluge can be tested without flooding chamber
Communications 4 station primary with backup sound powered phone
Atmospheric Conditioning Uses building hot and cold water
Gas Analysis Full time oxygen analysis
Patient Entertainment AM/FM cassette feed into chamber
Potable Water (optional) Hand washing sink in main lock for infections control as per JACO req'ts
Video Monitoring Provides full time monitoring of Chamber

Additional options are available for each system, which add features to increase functionality or remove unneeded features to reduce cost.

All Reimers Systems hyperbaric systems are designed to fully meet the current edition of all applicable codes including:

  • NFPA 99; "Health Care Facilities" Chapter 19, "Hyperbaric Chambers"
  • ASME PVHO-1; "Safety Standards for Pressure Vessels for Human Occupancy"
  • ASME Section VIII, Division I, "Unfired Pressure Vessels"