Reimers 8400D Series
Dual-Lock Multiplace Hyperbaric System

The 8400D series of multiplace chambers is designed for clinical hyperbaric therapy applications including those involving non-saturation decompression exposure. The dual lock design allows easy access into and out of the chamber while the chamber is at depth. The 8400B series is available in 6 patient (7 person) and 8 patient (9 person) sizes, with additional space in the entry lock. The entry lock of this chamber is designed for use anytime ingress and egress is necessary.

Features:

  • Exterior main lock is a large, easy to use 30"x 60" (76cm x 152cm) rectangular door.

  • Other doors are 30" (76cm) round doors

  • Six 10" diameter windows in main lock for maximum visibility of chamber interior. Two 10" windows in entry lock

  • Steel chairs are comfortable and removable to allow gurney access in main lock.

  • Attendant has folding jump seats that fold up out of the way.

  • Three external Canty lights in main lock, 2 in entry lock

 

Chamber Capacity:

  8406D 8408D
Patients, Seated, Main Lock 6 8
Patients, Seated, Entry Lock 3 3
Patients, Supine, Main Lock 2 2
Attendants, Seated, Main Lock 1 1
Attendants, Seated, Entry Lock 1 1

System Performance:

Pressure Capability  3.5 ATA (36 psig) 3.5 ATA (36 psig)
Pressure Capability, Optional 6.5 ATA (81 psig) 6.5 ATA (81 psig)
Pressurization Rate 6 psi/min 6 psi/min

Chamber Specifications:

Overall Length 17' 10" (544 cm) 19' 10" (605 cm)
Overall Width 7' 1" (216 cm) 7' 1" (216 cm)
Overall Height 7' 5" (226 cm) 7' 5" (226 cm)
Interior Length, Main Lock 10' 5" (318cm) 12' 5" (378cm)
Interior Width, Main Lock 7' 0" (213 cm) 7' 0" (213 cm)
Interior Length, Entry Lock 4' 8" (142cm) 4' 8" (142cm)
Interior Width, Entry Lock 7' 0" (213cm) 7' 0" (213cm)
Interior Ceiling Height 6' 3" (190 cm) 6' 3" (190 cm)
Overall Weight    

Hyperbaric System Includes:

Air Pressurization and Ventilation Air plant with dual oil-less compressors
  Automatic controls to maintain set pressure
  Manual pressurization and ventilation controls
Breathing Gas Patient stations support masks, hoods, and nebulizers
  Attendant stations support masks
  Automatically switches off oxygen if the fire system is activated
  High-pressure back-up oxygen and breathing air manifolds
Fire Suppression Deluge in mainlock and two hand lines
  Deluge and hand line are independent systems
  Bypass installed so deluge can be tested without flooding chamber
Communications 4 Station Primary with backup sound powered phone
Atmospheric Conditioning Uses building hot and cold water
Gas Analysis Full time oxygen analysis
Patient Entertainment AM/FM cassette feed into chamber
Potable Water (optional) Hand washing sink in main lock for infections control as per JACO req'ts
Video Monitoring Provides full time monitoring of Chamber

Additional options are available for each system, which add features to increase functionality or remove unneeded features to reduce cost.

All Reimers Systems hyperbaric systems are designed to fully meet the current edition of all applicable codes including:

  • NFPA 99; "Health Care Facilities" Chapter 19, "Hyperbaric Chambers"
  • ASME PVHO-1; "Safety Standards for Pressure Vessels for Human Occupancy"
  • ASME Section VIII, Division I, "Unfired Pressure Vessels"